Designed and Made in India for Indian & Global Healthcare Standards In-house R&D and manufacturing with strong service network Customizable for specialized departments (e.g., Oncology, Pediatrics) Affordable without compromising quality.
Smart Infusion, Smarter Care. The MaxFusion elevates infusion therapy with its intuitive interface, advanced safety features, and customizable delivery. Experience confident and comfortable care for patients, while simplifying workflows for clinicians. Discover intelligent and easy-to-use infusion.
The syringe pump supports a wide range of syringe sizes from 2 ml to 100 ml, accommodating major syringe brands through auto-recognition and manual calibration. Its advanced multi-mode operation, including Time/Rate, Body Weight, Dose, and Micro Infusion modes offers clinicians precise control over infusion parameters, ensuring safe and tailored drug delivery across diverse clinical scenarios.
Equipped with comprehensive audible and visual alarms, the syringe pump promptly alerts users to critical conditions such as occlusion, syringe dislodgment, low battery, end of infusion, KVO status, and system errors. Integrated anti-bolus technology further enhances patient safety by minimizing the risk of sudden fluid surges following occlusion release, crucial in high-risk infusion therapies.
Featuring a bright, Graphical LCD display, the syringe pump provides real-time visibility of infusion rate, volume, time, and pressure through an intuitive button-controlled interface. Its lightweight, space-saving design ensures easy mobility
Weight | 2.1kg | Size | 305 x 220 x 75mm |
Screen | 3.5-Inch LCD 128x64 Pixels | Brightness | 1-3 levels, adjustable |
Display | Infusion rate, VTBI, total volume, pressure limit & status, drug name, remaining time, status, battery status, size of syringe. | Mode | Rate Mode, Dose Mode, Target Mode, Auto Rate Mode. Micro mode Enabled Default. |
Accuracy | ≤ ±2% | Flow Rate | 0.1-2000ml/h |
Increament | 0.01ml, Use Navigation Keys to Easily Change the Flowrate value based on the position. | Preset Volume (VTBI) | 0.10 - 9999.99ml (increment: 0.01ml) |
Preset Time | 00:00:01-99:59:59, adjustable | Accumulated Volume | 0.1-9999.99 ml |
KVO | 0.01-9.99ml/h adjustable, step 0.01ml/h, default is 1.0ml/h | Purge | 0.01-2000ml/h |
Screen | 3.5-Inch LCD 128x64 Pixels | Bulse | 0.1-2000ml/h (automatic or manual) |
Occlusion Detection | 10 levels selectable: 0-2000mmHg, default is 500mmHg 2 units of pressure selectable: mmHg & psi default is mmHg |
Anti-Bolus | unexpected bolus reduced when the occlusion occurs |
Dose Rate units | ug/min, ug/h, mg/min, mg/h, ug/kg/min, ug/kg/h, mg/kg/min, mg/kg/h | Auto-Lock Time | 5 - 30 minutes selectable; ON/OFF switchable |
Drug Library | up to 20 drugs; | Event Log | up to 100 records |
Event (DERS) | automatic alarms when reaching dose limits | Syringe Compatibility | 2/5/10/20/50/60/100ml; automatic recognition of syringe size |
Operating Time | ≥ 10 hours at 5ml/h |
Charging Time | ≤ 12 hours to full capacity |
Interface | RS232,Wifi & USB |
Power Supply | voltage 100-240 V~ ; frequency 50/60Hz; current 0.4-0.14A |
Temperature | 5-40ºC for operating; -20-60 ºC for storage |
Relative Humidity | 15-95% for operating; 10-95% for storage |
Atmosphere Pressure | 57-106 kPa for operating; 50-106 kPa for storage |
Level of resistance to water and dust : IP23 |
Type | ≥Audible and visual alarm |
Sound Volume | 1-3 levels selectable, default is level 3 |
Alarm/Indication | Type | Decription |
---|---|---|
Stop Mode | Indication | The syringe pump is currently in a stopped state and Battery LED will also get Blink |
Run Mode | Indication | The syringe pump is actively running |
KVO Running | Indication | Keep Vein Open (KVO) mode is active to maintain a minimal flow. |
Low Battery | Warning | Low Battery Warning will be displayed in the LCD to indiate Battery is Low. Condition is battery percentage is less than 20% and Battery LED will also get Blink |
Occlusion | Alarm | When Blockages in vein or tube parth & Battery LED will also get Blink |
Target Reached | Alarm | The preset target volume has been delivered. |
Clamp Error | Alarm | "Indicates an error while loading the syringe or the clamp is not properly secured. And Battery LED will also get Blink" |
Disc Error | Alarm | "The syringe is not properly engaged or the drive disc is not in contact with the syringe plunger. And Battery LED will also get Blink" |
Driver Error | Alarm | The driver relay is stuck or pressed on while running condition and Battery LED will also get Blink |
Battery Critical | Alarm | "The battery voltage is critically low and requires immediate charging or power connection.This Condition battery percentage is less than 10% and Battery LED will also get Blink and shut down" |
Battery High Temp | Alarm | The battery temperature has risen beyond safe limits (45deg Celcius) and Battery LED will also get Blink. |
Int Err M1.1 | Alarm | Feedback from the motor indicates inconsistency or malfunction and Battery LED will also get Blink. |
Int Err M1.2 | Alarm | Unexpected reverse rotation detected during normal operation and Battery LED will also get Blink. |
Int Err M1.3 | Alarm | The motor is not running accurately as per the configured flow or steps and Battery LED will also get Blink. |
Int Err M1.4 | Alarm | Motor feedback noise detected, indicating instability and Battery LED will also get Blink. |
Board Disconneted | Alarm | "Communication lost between the Motherboard and the Display Board and Battery LED will also get Blink and shut down" |
System Error | Alarm | The load cell connector is disconnected or malfunctioning. |
© Copyright 2025 KIVIS Health Care LLP All Rights Reserved. Developed By Webindia Infotech